EasyManua.ls Logo

Alpha Moisture Systems dewTEC DS2000 - Appendices A - Product Specification

Alpha Moisture Systems dewTEC DS2000
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
Appendix A – Product Specification
1. Display
Type: Red/green, 7-segment, 5-digit primary display, 1-digit secondary display.
Height: 18mm (0.71inches) primary display, 7mm (0.3 inches) secondary display
Annunciators: Alarm 1 and Alarm 2 status
2. Sensor
Dewpoint Range: From –110°C to +20°C
Accuracy: ± 2°C over full range
Operating Temperature:
Storage Temperature:
Mechanical Connections:
Electrical Connection: Male coaxial plug
Approvals: CE, Atex -
3. Hygrometer Input
Sampling Rate: 10Hz
Resolution: 14-bits
Sensor short circuit detection: Detected within two seconds. All alarms become active
4. Transistor Outputs
Type: NPN open collector. Output 1 tied to Alarm 1. Output 2 tied to Alarm 2.
5. Relay Output 1
Contact Type: Single pole, double throw
Rating: 2A resistive, 120/240V, 50/60Hz
Lifetime: >500,000 operations at rated voltage/current
Isolation: Inherent
6. Relay Output 2
Contact Type: Single pole, double throw
Rating: 2A resistive, 120/240V, 50/60Hz
Lifetime: >500,000 operations at rated voltage/current
Isolation: Inherent
7. Linear (Re-Transmitted) Output
Ranges Available: 0-5V, 1-5V, 1-10V, 2-10V, 0-20mA and 4-20mA
Accuracy: ±0.5% max of display value
Load Impedance: 500 max for mA ranges, 500 min for voltage ranges
8. Serial Communications
Type: Serial asynchronous, UART to UART
Data Format: Open ASCII; one start bit, even parity, seven data bits and one stop bit.
Physical Layer: RS485
Maximum Number of zones 99
Baud Rate: Selectable – 1200, 2400, 4800 or 9600
9. Performance
Reference Conditions
Ambient Temperature: 20°C ± 2°C
Relative Humidity: 60 – 70%
Supply Voltage: 90 – 264V, 50Hz
Operating Conditions
Ambient Temperature: 0°C ± 55°C
Relative Humidity: 20 – 95% non-condensing
10. Environmental
Approvals: CE
EMC: Certified to EN61326
Notes: 1) For RF fields (10V/m 80% Am, 1kHz) the reading accuracy may be impaired
by up to –0.3% in the frequency band 87 to 109MHz.
2) For line-conducted disturbances induced by RF fields (10V 80% AM 1kHz).
The instrument is self recoverable in the frequency band 0.15-0.73 Hz
Safety: Complies with EN61010-1
Front Panel Seal: To IP66
Supply Voltage: 90 – 264V, 50/60Hz
Power Consumption: 7.5VA max
11. Physical
Dimensions: Height 48mm
Width 95mm
Depth 110mm without sensor connector.
Weight 0.21 kg max.
Mounting: Panel-mount; press-fit fixing. Panel cut out – 45mm x 92mm.
Terminals: Screw type
dewTEC DS2000 User’s Manual Page 22

Related product manuals