Veb- MN (Excell-MCDSe 400)#A(15-5-2007)(in).doc
7
conization electrodes in gynaecology) from 40-50W.
ADVICE
OPEN SKY SURGERY
For the Cut, use as follows:
Current PURE, from 40-50W, for a cut without the coagulating effect (it is not very common in surgery, where the coagulating cut is
preferred);
Current BLEND-1, from 40-50W, for a cut with a medium coagulating effect;
Current BLEND-2, from 40-50W, for a cut with a very strong superficial coagulating effect, spray type;
For the Coagulation, use as follows:
Current FULG FORCED, from 40-50W, for superficial coagulations or deep coagulations with the active electrode or the forceps. It is the
most suitable coagulation mode when a forceps or an electrode is used;
Current PIN POINT CONTACT, from 40-50W, for superficial coagulations or deep coagulations with the active electrode or the forceps, if
a softer superficial effect is desired, compared to the FULG FORCED coagulation;
Current SOFT, from 40-50W, for superficial or deep coagulations, which are softer than those of the PIN POINT CONTACT coagulation,
which is not suitable at all for the use with small electrodes with which it tends to cut;
Current SPRAY, from 40-50W, for a very strong superficial coagulation with the active electrode (without any contact with the tissues), and
a good deep coagulation with the forceps (it is the ideal coagulation mode for the operators who prefer the active electrode to coagulate).
LAPAROSCOPIC SURGERY
For the Cut, use as follows:
See above “OPEN SKY SURGERY” PURE, BLEND-1, BLEND-2. The current BLEND-2 is very efficacious when using the hooks or the
external part of the forceps/dissectors to cut with a strong coagulating effect.
For the Coagulation, use as follows:
See above “OPEN SKY SURGERY” FULG FORCED, PIN POINT CONTACT, SPRAY. The current SPRAY is very efficacious when
using the hooks or the external part of the forceps/dissectors to coagulate.