Troubleshooting the Purely Yours Breast Pump
Tocheckpumpsuctionfrommotor:
• Removetubingadapterfromtubingadapterportofpump.
• Putangerovertubingadapterport.
• Ifsuctionisfelt,pumpmotorisworkingproperly.
• Ifsuctionisnotfelt,unplugACadapter,waitafewseconds,plugbackintoelectricaloutlet.
ThePurelyYoursBreastPumphasaresetfeature.IfthePurelyYoursBreastPumpmotor
stopsrunning:
• UnplugACadaptercordfromelectricaloutlet.
• WaitafewsecondsandthenplugACadapterbackintooutlet.
Troubleshooting the HygieniKit Milk Collection System
Ifmilkisdrippingtooslowlyintobottle:
• SetSPEEDdialtoafastersetting.
• Replacesiliconediaphragm(s)andwhitevalve(s).
Ifpumpingcausesdiscomfort:
• SetpumptolowerSUCTIONsetting.
• Checkbreastanget.Alarger/smallerbreastangeorinsertmaybeneeded.
Ifsuctionislow:
• Checkthatwhitevalve(s)is/arepushedontobreastange(s).
• Checkwhitevalve(s)fortearsorstretches.Replaceifneeded.
• Ifsinglepumping,removeonetubeandcloseonesideofthetubingadapter.
• Checktubingadapterforcracks.Ifseen,replace.
• SetpumptohigherSUCTIONsetting.
Thewhitevalvesarevitalforpropersuction.Toworkwell,theymustbeintactandlookclosed
whenpumpisnotinuse.Toprolongwhitevalvelife,followcleaninginstructions.