SECTION 3 – MAINTENANCE 3-1
Section 3 - Maintenance
Cleaning
1. Drain all fluid from the viscometer.
2. For good measurement practice, the viscometer should be hand cleaned periodically to
remove fluid contamination. First, disconnect the encoder cable and unthread the large
coupling nut to remove the measurement head.
3. Carefully remove the measurement head.
Note: The measurement head has an O-ring seal and will be tight during removal.
4. Carefully wipe dry the inside of the viscometer and the bob.