EasyManua.ls Logo

Ametek 7270 - Page 202

Ametek 7270
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix E, ALPHABETICAL LISTING OF COMMANDS
E-10
according to the following table:
n Variable Range
0 X or X1 ±30000
1 Y or Y1 ±30000
2 MAG or MAG1 0 to 30000
3 PHA or PHA1 ±18000
4 ADC1 ±10000
5 ADC2 ±10000
6 ADC3 ±10000
7 ADC4 ±10000
8 C1 0 to 100000
9 C2 0 to 100000
10 0 zero
11 1 unity
12 FRQ 0 to 2000000000 (i.e. reference frequency in mHz)*
13 OSC 0 to 2000000000 (i.e. oscillator frequency in mHz)*
Dual modes only:-
14 X2 ±30000
15 Y2 ±30000
16 MAG2 0 to 30000
17 PHA2 ±18000
*Parameter n
5
is the only one that can be set to values of 12 or 13.
DEMOD2SRC [n] Second stage demodulator signal source
In the dual reference mode, the value of n sets the source of the signal for the second
stage demodulators according to the following table:
n Signal source
0 Main signal channel ADC (i.e. both demodulators are fed the same signal)
1 ADC1 - rear panel auxiliary ADC input
2 Demodulator 1 X channel output - this is the Tandem demodulation mode
ENBW[.] Equivalent noise bandwidth
In fixed point mode, reports the equivalent noise bandwidth of the output low-pass
filters at the current time constant setting in microhertz.
In floating point mode, reports the equivalent noise bandwidth of the output low-pass
filters at the current time constant setting in hertz.
EQU n Output result of equation #1 or equation #2
The value returned is the output of the user equation #1 (n = 1) or equation #2 (n =
2), where the equations are defined using the Output equations menu (see section
5.3.08). The possible range is ±2,147,483,647 (signed 32-bit integer).
EVENT [n] Event marker control
During a curve acquisition into the standard curve buffer, if bit 13 in the CBD
command has been asserted, the lock-in amplifier stores the value of the Event
variable at each sample point. This can be used as a marker indicating the point at
which an experimental parameter was changed. The EVENT command is used to set
this variable to any value between 0 and 32767.

Table of Contents

Related product manuals