ANALOG REMOTE CONTROL (Single XEL-P models only)
Non-isolated inputs and outputs to set voltage and current limit
Control input and
output scaling:
Rear panel control inputs (CV and CC) permit external 0V to 5V or 0 to
10V signals to set 0 to 100% of rated output voltage and current.
Set values of 0 to 100% of rated output voltage and current generate
0 to 5V signals at the rear panel V
out
and I
out
outputs.
These signals are referenced to the positive output.
Control input
accuracy:
Voltage: 0.3% ±10mV. Input impedance 100kΩ; protected to 60V.
Current: 0.5% ±5mA. Input impedance 64kΩ; protected to 60V.
Control output
accuracy:
Voltage: 0.3% ±10mV
Current: 0.5% ±5mA.
Output impedance: 125Ω; output short-circuit protected.
Remote Off: Rear panel socket allows a switch closure or logic low to turn output off.
DIGITAL INTERFACES (XEL-P models only)
Full digital remote control facilities are available through the RS232,USB, LAN and GPIB interfaces
Voltage Setting: Resolution 1mV, Accuracy ± (0·05% + 5mV)
Voltage Readback: Resolution 1mV, Accuracy ± (0·1% + 5mV)
Current Setting and
Current Readback:
Resolution 0.1mA; 0.01mA on 500mA range.
Accuracy ± (0·3% + 3mA) to 3A, ± (0·5% + 3mA) to 6A;
± (0·3% + 0·3mA) on 500mA range.
RS232: Standard 9-pin D-connector. Baud rate 9600.
GPIB (optional): Conforming with IEEE488.1 and IEEE488.2
USB: Standard USB 2.0 hardware connection. Operates as a virtual COM port.
LAN: Ethernet 100/10base-T hardware connection. LXI V1.2, Class C
compliant.
Remote Command
Processing Time:
Typically <25ms between receiving the command terminator for a step
voltage change at the instrument and the output voltage beginning to
change.
GENERAL
AC Input: 230V AC or 115V AC ± 10%, 50/60Hz. Installation Category II
Power Consumption: Single: 280VA max. Dual: 560VA max.
Operating Range: +5ºC to +40ºC, 20% to 80% RH
Storage Range:
−40ºC to + 70ºC
Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2.
Cooling: Intelligent variable-speed low noise fan assists convection. Over-
temperature trip shuts down output if internal temperatures exceed
predetermined thresholds.
Safety:
Complies with EN61010−1
EMC: Complies with EN61326
Size: Single: 107mm x 131mm (¼ rack 3U) x 288mm L (XEL-P: 343mm L),
excluding feet, knobs and terminals.
Dual: 214mm x 131mm (½ rack 3U) x 288mm L, excluding feet, knobs
and terminals.
Weight: Single: 4·5kg (XEL-P: 4.9kg). Dual: 8.5kg.
7