APPENDIX A
SCPI Command Reference
Table A-4 SCPI Commands for Calibration
Restore
Factory
Calibration
[:]CALibration[<channel>]:RESTore
Restores the calibration to the
factory defaults.
Restore
Default
Calibration
[:]CALibration[<channel>]:DEFault
Restores the default
calibration for all variables.
(Unit gain and zero offset).
Calibrate OVP
protection.
[:]CALibration[<channel>][:VOLTage] :P
ROTection[:OVER]
Calibrate the over voltage
protection.
<OVP Voltage Set in
Volts>
Enter Output
Voltage Data
[:]CALibration[<channel>]:OUTPut :VOLT
age
Sets the voltage output
calibration level.
Enter Output
Current Data
[:]CALibration[<channel>]:OUTPut :CURR
ent
Sets the current output
calibration level
Calibrate the
Voltage
controlled Non
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage]
Calibrate the voltage monitor
signal system.
? | <NR1> | <nothing
for interactive mode>
Calibrate the
Voltage
controlled
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage]:ISOLated
Calibrate the isolated voltage
monitor signal system.
? | <NR1> | <nothing
for interactive mode>