EasyManua.ls Logo

Ametek Sorensen XG 20-76 - Page 106

Ametek Sorensen XG 20-76
399 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-6 M370430-01 Rev F
Pin
Reference
Function
J1.13
VOL_MON
1
Voltage Monitor. The scale of this output is set by the APG voltage
input range selected (see Important note below).
J1.14
CUR_MON
1
Current Monitor. The scale of this output is set by the APG current
input range selected (see Important note below).
J1.15
EXT_CC_CV
External CC/CV. Indicates the state of the operate mode. When in
CV mode, logic high is output and when in CC mode, logic low is
output. Logic is TTL/CMOS compatible.
J1.16
CSH
Current Share. Used to hook up units for current sharing. Output
from master unit. Slave units should be left open.
J1.17
REF_I
Current Reference Input. Slave input from current share output on
the master unit.
J1.18
NC
1. Important: Non-isolated monitor lines are only valid in local control, digital remote and
non-isolated analog programming modes for the respective voltage and current control. If isolated
analog programming is used, the isolated monitor lines should be used as the non-isolated
monitoring lines will not accurately reflect the output state of the power supply.

Table of Contents

Related product manuals