D-0116493-B (OM015) PC850 Operating Manual
6. OPTIONS AVAILABLE TO SET-UP
6.1. GENERAL
The PC850 software stores the most recent test configuration and will initially use this to run any subsequent
tests. However, there are a number of options available for setting up, modifying and controlling a test.
These are summarised below - please refer to the operating window of the PC850 Software.
The “Options” menu selection at the top of the PC850 window opens a dialogue box, which provides access
to a number of options as follows:
Frequencies, which allows the operator to add or delete specific frequencies to or from the test
regime; note that 1kHz is always included
Computer test, which provides controls to allow:
o the selection of either 2 out of 3, or 3 out of 5 consistent responses to generate a valid
threshold
o a familiarisation run prior to the test
o a retest of the first frequency to be omitted (including this function can prove useful in
correlating test results)
Tone response, which provides controls to determine the action to be taken on error:
o the number of times a frequency is repeated (0, 1, 2 or 3 times) if an error in testing occurs
(for example, if there is no response from the patient)
o and then the action to be taken if the error continues (skip the frequency, stop the test or
pause the test)
Patient details, which may be ignored when used with Audibase
COM, which may be ignored for instruments with serial numbers covered by this user guide
Colour display, which may be selected to display the audiogram in colour
Colour print, which may be ignored when used with Audibase
Beep on finish, which sounds an audible warning on the PC at the end of the test