38
D-0116493-B (OM015) – PC850 Operating Manual
Guidance and manufacturer’s declaration – electromagnetic emissions
The PC850 Audiometer is intended for use in the electromagnetic environment specified below. The customer or
user of PC850 Audiometer should assure that it is used in such an environment.
Electromagnetic environment – guidance
The PC850 Audiometer uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause
interference in nearby electronic equipment.
The PC850 Audiometer is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity (1)
The PC850 Audiometer is intended for use in the electromagnetic environment specified below. The customer or
user of the PC850 Audiometer should assure that it is used in such an environment.
Electromagnetic
environment – guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%