16
SPECIFICATIONS
Display 3-5/6 digits 6000 counts
Sensing True RMS
Polarity Automatic
Update rate 5 per second nominal
Operating
temperature
14 °F to 122 °F (-10 °C to 50 °C)
Relative
humidity
Non condensing 10 °C, 10 °C to 30 °C 90%,
30 °C to 40 °C 75%, 40 °C to 50 °C 45%
Storage
temperature
-4 °F to 140 °F (-20 °C to 60 °C), < 80% R.H. (with battery
removed)
Pollution
degree
2
Operating
altitude
2000 m (storage below 12000 m)
Temperature
coefficient
nominal 0.10 x (specified accuracy)/
O
C @(0
O
C to 18
O
C or
28
O
C to 40
O
C), or otherwise specified
Transient
Protection
8.0 kV (1.2/50 μs surge)
Overload
protections
Current & Hz functions via jaws: 1000 A ac/dc rms at
< 400Hz
Voltage & 3-Phase Rotation functions via terminals:
1100 V dc / ac rms
Other functions via terminals: 1000 V ac/dc rms
E.M.C.
Meets EN61326-1:2006
DC A and DC+AC A Functions, in an RF field of 1V/m:
Total Accuracy = Specified Accuracy + 60 digits at around
200 MHz to 350 MHz
DC μA and Ohm Functions, in an RF field of 1V/m:
Total Accuracy = Specified Accuracy + 80 digits
Other Functions, in an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 20 digits
Agency
approval
P
Power supply Two 1.5V AA Size battery
Power
consumption
Typical 13mA for Current functions; 4.3mA for others
Low battery
indication
Approx. 2.85 V for Capacitance & Hz; approx. 2.5 V for
other functions
Auto power
off
Idle for 32 minutes
Auto power
off power
consumption
5μA typical