Operating Manual
novAA 800
Atomic Absorption Spectrometer

Table of Contents

Other manuals for Analytik Jena novAA 800

Questions and Answers

  • C
    Curtis WilsonAug 18, 2025
    How to fix if the flame does not ignite on Analytik Jena Laboratory Equipment?
    • C
      colemanmariahAug 18, 2025
      If the flame does not ignite in your Analytik Jena Laboratory Equipment, it might be due to an error with the gas supply. Ensure that both the air and fuel gas supplies are properly connected and functioning.
  • J
    Joel CarneyAug 21, 2025
    What to do if the cooling water flow is too low in Analytik Jena Laboratory Equipment?
    • T
      Troy MatthewsAug 21, 2025
      If your Analytik Jena Laboratory Equipment displays a 'Cooling water flow too low!' message, it indicates that the cooling water flow is insufficient. Check the cooling water level in the cooling unit and top it up. If the issue persists, it might be due to a partial obstruction of the cooling water channels, in this case you should also top up the cooling water. For obstructions, contact customer service.
  • M
    Mathew RobbinsAug 23, 2025
    Why is the line source signal or emission signal too small in Analytik Jena Laboratory Equipment?
    • S
      Stephanie WashingtonAug 23, 2025
      If the line source or emission signal is too small on your Analytik Jena Laboratory Equipment, it could be due to a weak HCL. Check the calibration state of the HCL and the atomizer. You can also try increasing the slit width and the lamp current or choosing a more energy-intensive HCL line.
  • S
    stephaniecharlesAug 26, 2025
    What causes big energy drift or energy fluctuations in Analytik Jena Laboratory Equipment?
    • T
      Timothy HernandezAug 26, 2025
      If you're experiencing significant energy drift or fluctuations with your Analytik Jena Laboratory Equipment, the HCL energy might be fluctuating too much. Start up the HCL and wait for 5 to 15 minutes before manually repeating an energy alignment and a peak search. Also, check the calibration state of the HCL and the atomizer. You can also try increasing the slit width and the lamp current, choosing a more energy-intensive HCL line, and checking the aging (blackening) of the HCL; replace it if needed.
  • S
    Steven SandersAug 28, 2025
    What to do if there is too little energy during the zero balance (AZ phase) in Analytik Jena Laboratory Equipment?
    • R
      Robert FieldsAug 28, 2025
      If there is too little energy during the zero balance (AZ) phase in your Analytik Jena Laboratory Equipment, it may be due to drift of the HCL energy level. Start up the HCL and wait for 5 to 15 minutes before manually repeating an energy alignment and a peak search. The AZ phase might also be disturbed by a fogged window or an unspecified ABS signal, so cleaning the windows may help. It could also be caused by the atomizer not being ideally aligned with the beam path, so check the calibration distance of the atomizer.
  • T
    Tina StevensAug 30, 2025
    Why is the background radiator signal too low in Analytik Jena Laboratory Equipment?
    • A
      Annette BryanAug 30, 2025
      If the background radiator signal is too low in your Analytik Jena Laboratory Equipment, it may be due to the energy of the D2 HCL being too low. Try increasing the slit width or selecting a HCL analysis line with a shorter wavelength. Also, check the alignment and aging (blackening) of the D2-HCL. Check the calibration distance of the atomizer. If necessary, do without the background compensation.
  • S
    Shane HarveySep 1, 2025
    What causes too much energy during the zero balance (AZ phase) in Analytik Jena Laboratory Equipment?
    • S
      stephaniekellerSep 1, 2025
      If there is too much energy during the zero balance (AZ phase) in your Analytik Jena Laboratory Equipment, it could be due to the drift of the HCL energy level being too high. Start up the HCL and wait for 5 to 15 minutes before manually repeating an energy alignment and a peak search. It may also be caused by an inappropriate moment in the measurement process for the AZ phase.
  • M
    Morgan GutierrezSep 3, 2025
    Why are energy drift or energy fluctuations in the background radiator too big in Analytik Jena Laboratory Equipment?
    • N
      Nicole SmithSep 4, 2025
      If the energy drift or energy fluctuations in the background radiator are too big in your Analytik Jena Laboratory Equipment, the fluctuations of the D2-HCL energy might be too significant. Try increasing the slit width or selecting an analysis line with a shorter wavelength. Also, check the alignment and aging (blackening) of the D2-HCL and the calibration distance of the atomizer.
  • C
    Christopher NelsonSep 6, 2025
    What causes baseline drift during the AZ phase (total absorption) in Analytik Jena Laboratory Equipment?
    • M
      millernicholasSep 6, 2025
      If you observe baseline drift during the AZ phase (total absorption) in your Analytik Jena Laboratory Equipment, it may be caused by a significant drift of the HCL energy level. To address this, start up the HCL and wait for 5 to 15 minutes before manually repeating an energy alignment and a peak search.
  • J
    jessicagreerSep 8, 2025
    What to do for a firmware update communications error on Analytik Jena Laboratory Equipment?
    • R
      Robert JonesSep 8, 2025
      If you encounter a firmware update communications error with your Analytik Jena Laboratory Equipment, it indicates that the firmware update has failed. You should repeat the firmware update. If the problem persists, inform service.

Summary

Safety Instructions

General Notes and Device Labeling

General safety advice, device symbols, and operating personnel requirements.

Safety Instructions: Transport and Installation

Provides safety guidelines for moving and setting up the device.

Safety Instructions for Operation

Covers essential safety measures during the device's operational use.

Behavior During Emergencies

Procedures to follow in case of hazardous situations or accidents.

Function and Setup

Installation and Initial Start-up

Loading and Adjusting Lamps

Procedures for installing, configuring, and aligning lamps for optimal performance.

Service and Maintenance

Troubleshooting

Troubleshooting According to Software Messages

Lists error codes and messages with corresponding causes and remedies.

Equipment Faults and Analytical Problems

Addresses common equipment issues and implausible analytical results.

Transport and Storage

Specification

Analytik Jena novAA 800 Specifications

General IconGeneral
BrandAnalytik Jena
ModelnovAA 800
CategoryLaboratory Equipment
LanguageEnglish