Anatel A643a - Anatel A643a Setup 71 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
7) Use the Up and Down Keys to specify one of the two TOC Print outputs.
8) Press Eto select the specified format.
• Continuous generates a printout at the conclusion of each analysis.
• Paper Saver generates a printout only when the difference in TOC values
between the current and the previous analysis exceeds a user-specified
percentage. If this format is chosen, the % Change in TOC must be specified.
A flashing cursor underscores the setting.
9) Use the Up and Down Keys to scroll the value until the desired percentage is
displayed.
10) Press Esc repeatedly to exit this function and return to the normal display.
5.6 The Purge Mode
In the Purge Mode, the Anatel A643a’s internal sample valve is open to permit sample water to
continuously flow through its measurement cell. This action provides a method of reporting the
conductivity (or resistivity) and temperature values on the Controller’s display and on local
printouts.
To place the Analyzer in the Purge Mode:
1) With the desired instrument selected in any View, press the Menu Key to display its
options.
2) Use the Up and Down Keys to specify Modes and press Enter.
3) Select Purge from the available selections and press Enter to enable the Purge Mode.
The Analyzer remains in this operational state until another Mode is chosen.
1 SENSOR NAME
X Continuous
Paper Saver:
Fig 5-35 : TOC Print Format
Fig 5-36 : Paper Saver Print Screen