5.6.1 Purge Printouts
Note:
Default: Timed / 1 Minute
As with the Auto TOC Mode, there are two potential Purge Print intervals: Timed generates a
purge printout at user-specified intervals; Paper Saver triggers a purge printout only when the
change in the water sample’s conductivity (or resistivity) level exceeds a userdefined
percentage in two successive analyses. The Purge Printout reports the following values:
Fig 5-37 : Selecting the Purge Mode
Table 5-3 : Purge Printout Reports
Sensor
The Model Number of the instrument.
SN
The Serial Number of the instrument.
ID
Its Channel ID Number.
Name
The user-entered name of the instrument.
Data
The Current Date
TIME
The time at which the measurement was taken.
COND UNCMP
The conductivity of the sample and whether it is temperature-uncompensated.
TEMP °C
The temperature of the sample in °C.