23. Cleaning and Disinfecting the Trolley
It is recommended that only CE marked cleaning products are used in the cleaning of
the trolley and the mattress. Cleaning and disinfection should be carried out by hand
only.
CAUTION:
Do not steam clean or jet wash this device.
Do not soak or immerse this device.
Do not use concentrated bleaching disinfectant solutions, organic solvent
or abrasive powders in the cleaning or disinfection of this product.
Dilute all disinfectants in accordance with the manufacturer’s guidelines.
Disinfectant products are corrosive in nature; failure to properly wash and
dry the product surface could leave a corrosive residue which may cause
damage to the product. Ensure the mattress is thoroughly dried before
refitting.
Clean the trolley and mattress with warm water and neutral detergent and dry the
surfaces thoroughly using a soft cloth. Suitable disinfectants are: quaternary
ammonium compounds, isopropyl alcohol & chlorine bleach; refer to the table below.
Apply disinfectant by cloth, spray or disinfectant wipe. Following disinfection, wash off
all surfaces with clean warm water and dry thoroughly using a soft cloth. Clean all
touch fastener attachments periodically with a soft brush, neutral detergent and
suitable disinfectant as stated. The product will be adversely affected and its life
expectancy reduced if the above cautions are not observed.
Compound
% Active
pH Range
Dwell Time
Chlorine
<10,000ppm (1%)
7-9
≤10 mins.
Chlorine is suitable, ensure the surfaces are rinsed and thoroughly dried as instructed.
Alcohol
70 (typical)
N/A
≤10 mins
Alcohol is suitable, subject to the following precaution; the surfaces must be rinsed
and thoroughly dried as instructed (if not the PU coating may swell and be vulnerable
to damage).
Hydrogen Peroxide
3-25%
5-9
≤5 mins.
Hydrogen Peroxide is conditionally suitable; the compound must be fully neutralized,
and the surfaces must be rinsed and thoroughly dried as instructed. Highly alkaline
peroxide pH≥10 is not suitable.
Quaternary Ammonium
3-15%
7-13
Varies
Quaternary Ammonium is conditionally suitable; the surfaces must be rinsed and
thoroughly dried as instructed. Quaternary Ammonium’s are typically too alkaline,
those with a pH≥10 are not suitable. Do not use QUAT wipes containing sodium
hydroxide.
24. Product Warranty
The product, when new, is guaranteed to be free from defects in materials and
workmanship and to perform in accordance with the manufacturer’s specification for a
period of one year from the date of invoice from Anetic Aid or their approved
distributor. Anetic Aid will repair or replace, at their discretion, any components found
to be defective or at variance with the manufacturer’s specification within this time at
no cost to the purchaser.
Protect your investment with a manufacturer backed service and
maintenance package; contact Anetic Aid for more details.
Warranty exclusions; the warranty does not provide cover for breakage or failure due to
tampering, misuse, neglect, accidents, modifications or shipping. The warranty is also
void if the product is not used in accordance with the manufacturer’s instructions or is
repaired during the warranty period by any persons other than Anetic Aid or its
appointed agent. No other expressed or implied warranty is given.