General Product Information
Document No. 992010. Issue 15 12.01.21
Page 18
Extended warranty; the warranty may be extended from the date of purchase, if the
product is maintained by Anetic Aid or its appointed distributor, commencing at the end
of the initial one year warranty period (quotations available upon request). Extended
warranty limitations; the extended warranty does not cover pressure care mattresses or
ancillary equipment (12 month warranty only applied).
For warranty, service and calibration, please contact Anetic Aid or their appointed
distributor.
25. Product Maintenance
The life expectancy of a QA4 Surgery Trolley is 10 years from date of introduction to
clinical use, dependant on the level of care and maintenance. The performance of this
device may reduce once the life expectancy has been reached and exceeded. It is
recommended that the trolley is serviced on an annual basis in accordance with the
manufacturer’s service schedule, and that the battery is replaced every 3 years.
Before use, ensure all trolley functions operate to their full range of movement and that
all components disengage, re-engage and lock correctly. Also visually inspect the trolley
for any loose or damaged parts and foreign bodies caught in the castors.
NOTE:
If the trolley is damaged or faulty it must be taken out of use with immediate
effect and the fault reported to Anetic Aid, your authorised dealer or
maintenance department. The trolley must not be used until the damage or
fault has been repaired.
CAUTION: In line with the MHRA document, Managing Medical Devices,
maintenance work should only be conducted by suitably trained personnel
following manufacturer’s guidelines.
26. Label Identification
The following list is a description of all the labels used on the trolley;
Product reference, serial number and date
of manufacture.
Maximum patient weight limit is 250kg
(551lbs) and the trolley safe working load
is 300kg (661lbs). The maximum load for
the head section is 25kg (55lbs) and the
maximum load for the leg section is 50kg
(110lbs).
Depress the brake pedal to brake all four
castors.