30. Using the Urology, Retrograde Pyelography, Gynaecology Extension
The purpose of the accessory is to increase patient c-arm access by enabling the patient
to be positioned further down the trolley, towards the foot end, and away from the
central column area; see fig.5, ‘C’ Arm Access. To detach the leg section, and attach the
extension (A), refer to Section 16. Using the Leg Section.
NOTE:
To maximise trolley stability, the castors (B) are best positioned projecting
away from the trolley, and underneath the extension, as shown in the
image below.
NOTE:
When transporting patients, traverse the trolley to the neutral position to
ensure maximum stability; see Section 10. Using the Patient Platform
Traverse.
CAUTION: To prevent the trolley becoming unstable, do not fit the
following accessories in combination;
Do not fit more than one retrograde extension.
Do not fit a foot end mattress extension in combination with a leg
section and retrograde section.