7.1 Status Commands
7-3
7
Remote Commands
:INSTrument:DEMUX:PTRansition?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
0 to 2112 Sum total of the bit of transition filter (DECIMAL)
Function Queries the condition of the transition filter (positive transition) of the
64Gbit/s DEMUX status.
Example > :INSTrument:DEMUX:PTRansition?
< 64
:INSTrument:DEMUX:NTRansition <numeric>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
0 to 2112 Sum total of the bit of transition filter (DECIMAL)
Available bit
64 (Bit 6) Delay Busy
2048 (Bit 11) Delay Calibration Require
Function Sets the transition filter (negative transition) of the 64Gbit/s DEMUX
status.
Example To set the transition filter (negative transition) of the 64Gbit/s DEMUX
status to 1:
> :INSTrument:DEMUX:NTRansition 64
:INSTrument:DEMUX:NTRansition?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
0 to 2112 Sum total of the bit of transition filter (DECIMAL)
Function Queries the transition filter (negative transition) of the 64Gbit/s DEMUX
status.
Example > :INSTrument:DEMUX:NTRansition?
< 64
:INSTrument:DEMUX:RESet
Function Initializes event at 64Gbit/s DEMUX status.
Example > :INSTrument:DEMUX:RESet