7.2.2 File menu commands
Table 7.2.2-1 File Menu Commands
No. Item Command
1 Open
:SYSTem:MMEMory:DEMux:RECall
2 Save
:SYSTem:MMEMory:DEMux:STORe
3 Print
:SYSTem:PRINt:DEMux
:SYSTem:MMEMory:DEMUX:RECall <file_name>
Parameter <file_name>=<STRING PROGRAM DATA>
"<drv>:\[<dir>]<file>"
<drv> = C,D,E,F
<dir>=<dir1>\<dir2>\... (Omitted for the root directory)
<file> = File name
Function Opens DEMUX setting data.
Example > :SYSTem:MMEMory:DEMUX:RECall "C:\Test\example"
:SYSTem:MMEMory:DEMUX:STORe<file_name>,<data_type>,<file_type>
Parameter <file_name>=<STRING PROGRAM DATA>
"<drv>:\[<dir>]<file>"
<drv> = C,D,E,F
<dir>=<dir1>\<dir2>\... (Omitted for the root directory)
<file> = File Name
<data_type>=<CHARACTER PROGRAM DATA>
UDX 56G DEMUX Setup
UER 56G DEMUX Error Alarm Result
<file_type>=<CHARACTER PROGRAM DATA>
BIN Binary file
CSV CSV file
TXT Text file
Function Saves DEMUX setting data.
Note:
The settings will not be read from the saved file if the file name is
changed.
Example To specify save destination for all setting files and save them with a
comment and measurement result data:
> :SYSTem:MMEMory:DEMUX:STORe "C:\Test\example",UMX,TXT