7.4 Automatic Measurement Commands
7-47
7
Remote Commands
:SENSe:MEASure:EMARgin:MARGin:RESolution <type>
Parameter <type>=<CHARACTER PROGRAM DATA>
FINE Fine mode
COARse Coarse mode
Function Sets the accuracy of the error ratio for the Eye Margin measurement.
Example To set the accuracy of the error ratio for the Eye Margin measurement to the
Fine mode:
> :SENSe:MEASure:EMARgin:MARGin:RESolution FINE
Compatibility Compatible with the MP1632C.
:SENSe:MEASure:EMARgin:MARGin:RESolution?
Response <type>=<CHARACTER RESPONSE DATA>
FINE, COAR
Function Queries the accuracy of the error ratio for the Eye Margin measurement.
Example > :SENSe:MEASure:EMARgin:MARGin:RESolution?
< FINE
Compatibility Compatible with the MP1632C.
:SENSe:MEASure:EMARgin:MARGin:ASEarch <boolean>[,<mode>]
Parameter <boolean>=<BOOLEAN PROGRAM DATA>
ON or 1 Auto Search ON
OFF or 0 Auto Search OFF
<mode>=<DECIMAL NUMERIC PROGRAM DATA>
1 Coarse
2 Fine
Note:
<mode> can be omitted. Coarse is specified when omitted.
Function Selects whether to perform Auto Search upon start of the Eye Margin
measurement.
Example To set Auto Search upon start of the Eye Margin measurement to ON:
> :SENSe:MEASure:EMARgin:MARGin:ASEarch ON
Compatibility Incompatible with existing models.