Chapter 7 Remote Commands
7-96
:SENSe:MEASure:BATHtub:CALRange?
Response <upper>,<lower>=<CHARACTER RESONSE DATA>
Function Queries the error threshold range during Bathtub measurement
recalculation.
Example > :SENSe:MEASure:BATHtub:CALRange?
< E_5,E_10
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:ASEarch
Parameter <item>=<CHARACTER PROGRAM DATA>
OFF Auto Search OFF
PHASe Phase auto search (Coarse)
PTHReshold Phase & threshold auto search (Coarse)
FPHase Phase auto search (Fine)
FPTHreshold Phase & Threshold auto search (Fine)
Function Sets execution of auto search during Bathtub measurement.
Example To set the phase auto search ON:
> :SENSe:MEASure:BATHtub:ASEarch PHASe
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:ASEarch?
Response <item>=<CHARACTER RESPONSE DATA>
OFF, PHAS, PTHR, FPH, FPTH
Function Queries the execution of auto search during Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:ASEarch?
< PHAS
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:CURRent <boolean>
Parameter <boolean>=<BOOLEAN PROGRAM DATA>
0 or OFF OFF: Updates at measurement termination.
1 or ON ON: Updates every second.
Function Sets the graph display update period during Bathtub measurement.
Example To set the graph display update timing to every second:
> :SENSe:MEASure:BATHtub:CURRent 1
Compatibility Incompatible with existing models.