Anritsu Site Master S331D - Power Meter Relative Mode On;Off - Control Byte #41 (29 H) (Option 29 Only); Power Meter Offset Mode On;Off - Control Byte #42 (2 Ah) (Option 29 Only); Power Meter Zero Mode On;Off - Control Byte #43 (2 Bh) (Option 29 Only); Power Meter RMS Averaging On;Off - Control Byte #44 (2 Ch) (Option 29 Only)

Anritsu Site Master S331D
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Power Meter Relative Mode On/Off Control Byte #41 (29h) (Option 29 only)
Description: Enable or disable Power Meter Relative Mode.
Bytes to Follow: 1 byte
1) Relative Mode State
00h = Off
01h = On w/ trigger (use the current power level as a reference power level)
Site Master Returns: 1 byte
1) 255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid parameter
238 (EEh) Time-out Error
Power Meter Offset Mode On/Off Control Byte #42 (2Ah) (Option 29 only)
Description: Enable or disable Power Meter Offset Mode.
Bytes to Follow: 5 bytes
1) On/Off (01h = On, 00h = Off)
2 - 5) Offset Power level in dB (Multiplied by 1000)
Site Master Returns: 1 byte
1) 255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid parameter
238 (EEh) Time-out Error
Note:If you turn the Offset mode off, you must still send the other bytes. Bytes2-5will be ignored.
Power Meter Zero Mode On/Off Control Byte #43 (2Bh) (Option 29 only)
Description: Enable or disable Power Meter Zeroing Mode.
Bytes to Follow: 1 byte
1) Zero Mode Status
00h = Off
01h = On with trigger (current power level is referenced as -80 dBm)
Site Master Returns: 1 byte
1) 255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid status
238 (EEh) Time-out Error
Power Meter RMS Averaging On/Off Control Byte #44 (2Ch) (Option 29 only)
Description: Set Power Meter RMS Averaging. Enabling to one of three different levels, or off.
Bytes to Follow: 1 byte
1) RMS Averaging State
00h = Off
01h = On (Low) with trigger (current power level is referenced as -80 dBm)
02h = On (Medium)
03h = On (High)
Site Master Returns: 1 byte