A3 PLUS - A5 - A6 PLUS - L6
SURGICAL SINGLE CART
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the foot board under
the dental chair has been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the dental chair
backrest has been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
Any movement request has been made during a disinfection
cycle.
Wait for the end of the disinfection cycle, then put the instruments
back in place to make movements available again.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the bowl has been
triggered.
Manually move the bowl out of the interference area.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the dental chair seat
has been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the side delivery arm
has been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
An automatic movement request A/B/C/D has been made with
one extracted instrument.
Put instrument back in place in the rest position to make
automatic movement programmes available again.
A movement request has been made with one extracted
instrument
To move the dental chair manually, disable the instrument. To
activate automatic movements, disable the instrument and put it
back in place.
A movement request has been made with dental chair lock option
enabled.
Remove the dental chair lock using the suitable button.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the backrest cover has
been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the footrest has been
triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the leg support has
been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
The bowl movement request cannot be implemented since the
dental chair is in the interference position.
Lower the dental chair in order to move it out of the interference
position, then request the bowl movement.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the bowl has been
triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
A movement has been interrupted or a movement request cannot
be implemented since the thermal cut-out of the dental chair
lifting motor has been triggered.
Wait for the motor to cool down.
If the warning message is not included in the list, please contact the Technical Service department for further details.