10. WARNING/ERROR MESSAGES
Two types of notices can be displayed on the console display: warning messages (Wxxx) and error messages (Exxx).
Warning messages (Wxxx) indicate situations that require user intervention.
Error messages (Exxx) indicate situations that cannot be fixed by the user and require the intervention of the Technical Service.
When contacting the Technical Service, please communicate the error number displayed.
End movements, lay down instruments.
The disinfection cycle preparation page opens with the
instruments not in rest position.
To start a disinfection cycle, all disinfection instruments must be
in a rest position.
BIOSTER start request performed without selecting any
instrument or cup.
Select the instrument, cup or suction tubes, then activate the
disinfection cycle.
The system has stopped the disinfection cycle after detecting
instruments changed during the cycle.
To restart the disinfection cycle after a lockout use the same
instruments extracted during the stopped cycle.
BIOSTER start request performed without any extracted
instrument or selected cup.
Select the instrument or cup, then start the disinfection cycle.
A user has requested to skip the disinfection pause.
At the end of the cycle, a warning is issued if the pause has not
been made.
Dentist’s board water cocks closed.
Check that the dentist’s board water cocks are open, if the
problem persists, contact technical service department.
Emptying is not possible since bypass is not selected.
Select mains water supply.
It is impossible to fill W.H.E. system internal tank
Open main water supply, if the problem persists, contact the
Technical Service department.
An instrument has been extracted with active foot control.
Release the foot control, then activate the instrument.
Instruments extracted at start-up.
Put the instruments back in place.
Check connections, if the problem persists, contact the technical
service department.
A Flushing request has been made with no extracted instrument.
Repeat the Flushing cycle request with instrument extracted.
A Flushing request has been made S.H.S. tank in reserve.
Fill the H
2
O tank and repeat the Flushing request. If the problem
persists, contact technical service department.
Flushing Ended with error.
Repeat the Flushing cycle.
Flushing is not possible since bypass with long cycle is not
selected.
Repeat the Flushing cycle request with mains water selected.
Suction tubes extracted at start-up.
Put suction tubes back in place.
Instruments extracted at start-up.
Put instruments back in place.
A movement has been interrupted or a movement request cannot
be implemented since the safety device of the assistant's board
arm has been triggered.
Remove the obstacle; if the problem persists, contact the
technical service department.
Suction tube washing with bottle not allowed
Deselect supply from bottle and repeat the suction tube washing
cycle request with mains water selected.
The suction tube washing has been intentionally interrupted
again.
During the washing cycle, suction tubes do not suck.
Check that suction tube terminals are open or that filters are
clean; if the problem persists, contact technical service
department
Suction tube washing successfully completed.
Water supply change requested during BIOSTER or Flushing
cycle.
End BIOSTER or Flushing cycle.
Scheduled maintenance is overdue.
Contact technical service department.
Wireless foot control battery is flat.
Connect the power cable to the foot control.
Wireless foot control battery is charged.
Disconnect the power cable from the foot control.
Wireless foot control not connected.
Activate the wireless foot control for at least 1 second.