7.2.2. MANUAL DISINFECTION CYCLE WITH S.H.S. SYSTEM
The S.H.S. system allows performing a manual disinfection cycle of the
water ducts of all the instruments on the dentist's board and of the
syringe located on the assistant's board.
To perform the disinfection cycle, proceed as follows:
1 Preparing the disinfectant solution:
Pour undiluted PEROXY Ag+ (or 3% oxygenated water) into the
tank marked with an orange band (supplied).
2 Disinfectant injection phase:
Replace the independent supply tank (a) (bottle marked with a
grey band) with the disinfectant tank (bottle marked with an orange
band) (see paragraph 7.2.1).
Make sure that there is a quantity of liquid equal to at
least 300 cc.
Make sure the spray cocks (d) found towards the bottom of the
board, are fully open (if they are not, either very little or no water at
all will flow out).
Press the button to fill the water in the cup for 5 (five) consecutive
times, thus filling 5 cups of water.
This step is very important because it allows you to ensure that all
the liquid in the pipes is replaced with the disinfectant.
Remove the instruments one at a time and allow water to flow for
at least 2 minutes using the CHIPWATER foot control for dynamic
instruments (see section 5.2.) and the water button for syringes.
Now the ducts contain disinfectant liquid.
Put the instruments back in place.
3 Disinfectant contact time:
PEROXY Ag+ (or 3% oxygenated water) must be left to dwell in the ducts for at least 10 minutes, but not more than 30 minutes.
4 Duct rinsing phase:
Replace the bottle of disinfectant (orange band) with the bottle of normal independent supply (grey band).
As in the previous phase, remove the instruments one at a time and allow water to flow for at least 2 minutes using the CHIP-WATER foot control
for dynamic instruments (see section 5.2.) and the water button for syringes.
Press the button to fill the water in the cup for five consecutive times, thus filling five cups of water.
This step is very important because it allows you to ensure that all the disinfectant fluid in the pipes is replaced with the independent water supply
liquid.
IMPORTANT at the end of the disinfection cycle, always replace the tank containing the disinfectant (tank with orange band) with
the normal one (tank with grey band).
NEVER USE A DISINFECTANT TO IRRIGATE THE PATIENT'S ORAL CAVITY.
PEROXY Ag+ storage.
For proper PEROXY Ag+ storage, follow the manufacturer’s instructions given on the package.
It is important to keep the package tightly closed and stored in a cool place at a temperature not exceeding 25°C.
Never leave PEROXY Ag+ or oxygenated water in the tank (a) for more than one month.
If you are going to be absent from the surgery for long periods of time (holidays), completely empty the tank (a) before leaving.