Symbol Parameter Setting Contents Content
Factory
Default
P1 Temperature Unit ˚C – ˚F ˚F
P2 Automatic Hold 5-20 seconds – O O
P3 Automatic Backlight O 1-8 minutes – O 1
P4 Automatic Power O 10-20 minutes – O 10
P5 pH Buer Series Selection USA – NIST USA
P6 pH Resolution 0.1 – 0.01 0.01
P7 pH Calibration Reminder
H-hours D-Days
(set up in ZenTest App)
/
P8 Back to factory default No – Yes No
TypeofContamination CleaningAgent SoakingTime
Alkaline coatings
Apera Electrode Cleaning
Solution (SKU: AI1166)
5 – 10 minutes
Acidic coatings 0.1M NaOH solution 5 – 10 minutes
Sulde-containing
substances
Thiourea
30– 60 minutes, leave until
junction discoloration
disappears.
Unknown substances
Apera Electrode Cleaning
Solution (SKU: AI1166)
30 – 60 minutes
Use a soft brush to help thoroughly clean o the contaminants. After the cleaning
procedure, the pH electrode should always be soaked in 3M KCl solution for 12-24 hours
for rehydration. A re-calibration is also necessary before a new pH measurement.