7. Technical Specifications
Item
Specificatio n s
Mode of Operation
Continuous
Pressure Range
4 20 cmH
2
O (adjustable in 0.5 cmH
2
O increment)
Ramp Time
0 45 minutes (adjustable in 5-minute increment)
Ramp Starting Pressure
4 to therapy pressure-1
Altitude Compensation
(Manual setting)
Level 1~8 for 0 ~ 8000 ft when the pressure set at 4-18 cmH
2
O.
Level 1~5 for 0 ~ 5000 ft when the pressure set at 18.5-20 cmH
2
O
Dimensions (W x D x H)
14.5 x 13.0 x 10.0 cm or 5.7” x 5.1” x 3.9”
Weight
Approximately 800 g or 1.76 lb
Sound Level
28 dBA at 10 cmH
2
O, 1 meter distance
Power Input
AC100-240V, 50/60 Hz, 0.5-0.3A
Environment
Temperature
Operating: +5° C to +35° C (+41° F to +95° F)
Storage/Transport: -15º C to +50º C (+5° F to +122° F)
Humidity
Operating: 15%RH to 95%RH non-condensing
Storage/Transport: 10%RH to 90%RH non-condensing
Air Tubing
Flexible plastic, 1.8m (approx.)
Classification:
Class II
Type BF, Applied Parts Nasal Mask.
Not suitable for use in the presence of a flammable anesthetic mixture.
IP21: Protected against solid foreign objects of 12.5 mm and greater;
Protected against vertically falling water drops.
Continuous operation.Continuous operation.
Note: The manufacturer reserves the right to modify the specification without notice.