Page #43 background imageLoading...
Page #43 background image
128 0X80
Relay 2 PH high trigger
value
2 READ/WRITE 0—1400 1000
1001
129 0X81
Relay 2 PH low trigger
value
2 READ/WRITE 0—1400 400
1001
130 0X82
Relay 2 ORP high
trigger value
2 READ/WRITE -2000—+2000 1000
11
131 0X83
Relay 2 ORP low
trigger value
2 READ/WRITE -2000—+2000 -1000
11
132 0X84
Relay 2 PH hysteresis
value
2 READ/WRITE 0—1400 50
1001
133 0X85
Relay 2 ORP hysteresis
value
2 READ/WRITE -2000—+2000 100
11
134 0X86
PH-20MA
corresponding value
2 READ/WRITE 0-1400 1400
1001
135 0X87
PH-4MA
corresponding value
2 READ/WRITE 0-1400 0
1001
136 0X88
ORP-20MA
corresponding value
2 READ/WRITE -2000—+2000 2000
11
137 0X89
ORP-4MA
corresponding value
2 READ/WRITE -2000—+2000 -2000
11
138 0X8A
-20MA
corresponding value
2 READ/WRITE
100
1000
101