14
EN
• When the user ventilates, the bar graph shows in real time the quantity of air/oxygen that must be insufflated
based on ERC and AHA guidelines.
▶ Display in continuous mode:
o Display in 30:2 and 15:2 mode:
• In continuous resuscitation mode, EOlife X generates a visual countdown to show the user when to ventilate
to deliver the ‘target frequency’ breaths per minute. When the green lung is displayed, the user must
ventilate.
In 30:2 mode, the countdown function is disabled as the person ventilating must wait for 30 chest
compressions before delivering 2 ventilations.
In 15:2 mode, the countdown function is disabled as the person ventilating must wait for 15 chest
compressions before delivering 2 ventilations.
In this mode the ventilation frequency depends on the person (or machine) performing the chest
compressions.
The green lung is shown permanently to indicate that the device is waiting to detect a ventilation to display
the bar graph.
Orange:
Volume < Ideal volume.
Green:
Ideal volume.
Red:
Volume > Ideal volume.
8. Real-time indication of how to ventilate:
▶ A bar graph shows whether the volume insufflated/exhaled is adequate:

Table of Contents