23
EN
IX. Technical data
1. Technical data related to the device
Specifications EOlife X
Dimensions (W x H x D)
130 mm x 75 mm x 30 mm
(5.11 inches x 2.95 inches x 1.18 inches)
Weight 170g ± 5 g (5.3 oz ± 0.011 oz)
Operating conditions
• Temperature between 0°C and 40°C (32°F and 104°F)
• Relative humidity between 15% and 95% (non-condensing)
• Atmospheric pressure range of 620 hPa (altitude of 4000 m) to 1,060 hPa
(altitude of -500 m)
Storage/transport conditions
• Temperature between -10°C and 35°C (-14°F and 95°F)
• Relative humidity between 10% and 90% (non-condensing)
Lifetime 5 years
Run time 5 hours
IP rating for protection against
solids, dust and ingress of water
IP44 (in use configuration, i.e., EOlife X, its battery and FlowSense X connected)
Electromagnetic compatibility
(EMC) according to EN 60601-
1-2
The control parameters and threshold values can be obtained from the manufacturer.
BLE module frequency range
Signal strength
Certification
2,4 GHz.
Max. 4 dBm.
Directive 2014/53/EU, FCC, IC, MIC, KC.
Screen
2.4 inches.
Resolution 320 x 240 pixels.
Measurement accuracy
• Vi (volume insufflated): ± 8.8% of the actual value measured under normal conditions
of use
• Vt (tidal volume) without leakage: ± 8.8% of the actual value measured under normal
conditions of use
• Freq (ventilation frequency): ± 1 cycle per minute