p3
Protection:
Front face: IP 65
Housing: IP20
Terminals: IP 20
Housing:
Self-extinguishing case of
black UL 94 V0 ABS.
2. SPACE REQUIREMENTS
Dimensions of the case: (with terminals)
96 x 48 x 124 mm
Mounting on panel
cut out 44 x 91 mm
external
seal
Mounting panel
max. thickness 30
case
tightenings
terminals
Plug off connectors on rear face for
screwed connectings (2.5mm², flexible
or rigid)
Display: ±10 000 points (14 mm)
Electroluminescent red (green optional)
2 alarm leds
Features of the outputs
Types of
OUTPUTS
1 analog insulated
Active
current
or Voltage
A1
2 inverting relays
R
Current: direct or reversed 0-20mA
Load impedance ≤ Lr 600Ω
Voltage: direct or reversed 0-10V
Load impedance ≥ Lr 500kΩ
2 setpoints per relay
configurable on the whole MR.
Hysteresis programmable from 0 to 100%.
Time delay programmable from 0 to 25 sec.
(8A/250 VAC on resistive load)
Features
Code
A3
General features
Galvanic partition:
2.5 kV eff. - 50 Hz - 1 min between supply, inputs, outputs
• Standard sampling time: 100 ms
• Common mode rejection rate: 130 dB
Serial mode rejection rate: 40 dB 50/60 Hz
• Zero drift compensation and self-calibration
• Conform with standards IEC 61000-6-4 on rejections and IEC 61000-6-2 immunity
(in industrial environment) IEC 61000-4-2 level 3, IEC 61000-4-3 level 3,
IEC 61000-4-4 level 4, IEC 61000-4-6 level 3.
CE marking according to the directive EMC 89-336.
20 to 270 VAC - 50/60/400 Hz and
20 to 300 VDC
Max. operating range
3 W max.
5.5 VA max.
Power draw
Power supply: