What to do if Arjo Auralis shows LOW PRESSURE and WAIT?
- WWesley HernandezJul 27, 2025
If your Arjo Auralis SXPS indicates 'LOW PRESSURE and WAIT', it could be due to several reasons. First, check for leaks in the Auralis mattress or VHDW cushion. Ensure the CRP rapid deflation unit on the mattress is closed, and the valves on the seat cushion are also closed. Verify that the tube set connector is correctly connected to the Auralis SXPS; you should hear an audible click when it's properly connected. Both indicators should turn on when operating pressure is reached.