Autorm mode
Activate: Press and
hold the button.
Comfort indicators ash
and mattress inates to
Autorm.
Extend: Press button to
extend 5 minutes. Max.
30 minutes.
If not extended, pump
goes back to previous
mode.
Deactivate: Press and
hold the button or select
another mode.
Pressure reached –
Autorm and Control
comfort indicators turn to
constant.
CPR rapid deation unit
Activate: Press both CPR release
buttons simultaneously.
Reset: Press down the open part
until it clicks into closed position.
Make sure the front part is open.
The air evacuates rapidly.
2s
Alternating mode, default setting
Activate:
Press button, the Alternating mode indicator is shown.
Adjacent cells inate and deate on a 10 minute cycle.
15 min
Start the Auralis system
When the pump is connected to a power
source it runs a self check:
Beep 10 seconds 2 beeps ready to use.
Press and hold Run/Standby button.
The following indicators are shown:
Run/Standby
Wait indicator
Alternating mode button
Comfort control indicator
Battery charge indicator
Green indicators on the front of the pump
Once the Wait indicator turns
o󰀨, place bed sheet on mattress.
Make sure that the CPR rapid
deation unit is clearly visible.
Place patient on the mattress or
seat cushion.
Alternating mode is set by default.
2
1
1
1
2
2
Auralis
Deactivate:
Select Reactive CLP or Autorm mode.
If no further therapy is required, remove patient from
the surface and turn o󰀨 the pump.