Questions and Answers

  • M
    Michael SullivanAug 12, 2025
    What to do if the Arjo Enterprise 9000X Medical Equipment bed is beeping?
    • L
      lisa49Aug 12, 2025
      If your Arjo bed is beeping, it might be running on its backup battery. Ensure the power cord is properly plugged into a working electrical outlet. Also, check the fuse in the mains plug, if applicable.
  • M
    Michelle GardnerAug 16, 2025
    Why is my Arjo Enterprise 9000X Medical Equipment bed difficult to maneuver?
    • W
      wtaylorAug 16, 2025
      If your Arjo bed is hard to move, check the brake pedals. If they are in the “steer” position, place them in the “free” position to allow easier movement.
  • J
    John WillisAug 19, 2025
    What to do if the Arjo Enterprise 9000X mattress platform cannot be lowered?
    • L
      Lauren EllisAug 19, 2025
      If the mattress platform on your Arjo bed won't lower, it could be due to a height control software error. Try raising the mattress platform to its maximum height to reset the software.
  • L
    Larry SmithAug 23, 2025
    What should I do if the Arjo Enterprise 9000X displays error code E300?
    • C
      Caleb MyersAug 23, 2025
      If your Arjo bed shows error code E300, it means a control button was pressed for too long. Remove pressure from all control buttons. If the error persists, you should contact an Arjo approved service engineer.
  • J
    jamesparrishAug 27, 2025
    What does error code AES mean on Arjo Enterprise 9000X when the mattress platform cannot be lowered?
    • P
      parkerrobinAug 27, 2025
      Error code AES, along with the inability to lower the mattress platform on your Arjo bed, indicates that the Anti-Entrapment system has been activated. Check and remove any obstructions from underneath the mattress platform. If the bed still won't lower, contact an Arjo approved service engineer.
  • K
    Kelly WalkerAug 30, 2025
    Why are one or more functions inoperative on my Arjo Medical Equipment bed?
    • T
      Tyler ThompsonAug 31, 2025
      If some functions aren't working on your Arjo bed, they might be locked on the ACP (Auxiliary Control Panel). Unlock the functions on the ACP to restore their operation.
  • M
    Monica RichardsonSep 3, 2025
    Why does the alarm sound when the patient moves on the Arjo Enterprise 9000X Medical Equipment bed?
    • H
      hayden43Sep 3, 2025
      If the alarm goes off when the patient moves on the Arjo bed, the movement detection threshold might be set too low. Try increasing the VariZone threshold setting.
  • J
    Justin RodriguezSep 7, 2025
    Why are there errors in the indicated patient weight on my Arjo Enterprise 9000X Medical Equipment?
    • K
      Kimberly GordonSep 7, 2025
      If you're seeing errors in the patient's weight on the Arjo bed, it could be due to an obstruction of the mattress platform. Check that the mattress platform isn't touching furniture, curtains, cables, etc. Also, ensure that weight hasn't been added to the bed without using the auto-compensation function.

Summary

General Warnings

Clinical Applications

Contra-indications and Safe Working Load

Lists situations where the bed is unsuitable and its maximum load capacity.

Installation

Operation

Brakes and Steering Functions

How to use the brake and steering functions for bed mobility.

Split Side Rail Operation

How to safely raise and lower the split side rails.

Manual CPR Backrest Release

Procedure for emergency backrest lowering to facilitate CPR.

Attendant Control Panel (ACP)

Overview of the caregiver control panel and its functions.

Advanced Features

Patient Weighing System Controls

Controls for the bed's integrated patient weighing system.

Weighing System Accuracy Precautions

Factors affecting weighing accuracy and precautions to ensure correct measurements.

Weighing System Initialisation

Steps to properly initialize the weighing system before patient use.

Weighing and Auto Compensation Procedures

How to weigh patients and use auto compensation for weight adjustments.

Weighing System Errors and Angle Display

Common error codes and how angles are displayed on the weighing system.

VariZone Patient Movement Detection

Overview of the patient movement detection system and its sensitivity.

VariZone Operation and Controls

How to activate, adjust, and deactivate the VariZone movement detection.

Anti-Entrapment System (Optional)

Description and function of the anti-entrapment system sensors.

Product Care

Decontamination Procedures

Instructions for cleaning and disinfecting the bed.

Preventive Maintenance (Caregiver Checks)

Daily and weekly checks to be performed by caregivers.

Troubleshooting Common Issues

Table of common symptoms, possible causes, and corrective actions.

Troubleshooting Error Codes

Further common error codes and their corresponding solutions.

Warranty and Service Information

Arjo Enterprise 9000X Specifications

General IconGeneral
BrandArjo
ModelEnterprise 9000X
CategoryMedical Equipment
LanguageEnglish

Related product manuals