8. Technical Data
General
Safe working load 250kg
Maximum patient weight 185kg
Product weight (approx.) 180kg
Audible noise 50dB approx.
Operating conditions
Temperature 10°C to 40°C
(50°F to 104°F)
Relative humidity 20% to 90% at 30°C, non-condensing
Atmospheric pressure 700hPa to 1060hPa
Electrical data
Power input
3A max. at 230V a.c. 50/60Hz
3A max. at 230V a.c. 60Hz (KSA)
5.8A max. at 120V a.c. 60Hz
6.8 max. at 100V a.c. 50/60Hz
Duty cycle 10% (2 min. on, 18 min. off)
Safety standards USA/Canada
EN/IEC 60601-1:2005 AMD1:2012
ANSI/AAMI ES60601-1 (2005)
AMD 1 (2012)
CAN/CSA-C22.2 No. 60601-1:14
IEC60601-2-52:2015
Electric shock protection Class I Type B
EMC Complies with IEC 60601-1-2:2014
Potential equalisation terminal
Complies with EN/IEC 60601-1:2005
AMD1:2012
Liquid ingress protection IPX4
Backup battery
2 x 12V series connected, sealed,
rechargeable lead/acid gel, 1.3Ah
Patient weighing system
Minimum verification interval (scale division) e 500g
Minimum capacity 10kg
Maximum capacity 250kg
Approvals
EC type approval no. UK3074. Complies
with 2014/31/EU Class III.