10
2.4 Display options
A service technician can set the information to be shown in the
display.
2.4.1 Displayed while a program is running
The display can be set to show one of the following alternatives
while a program is running:
• Temperature
• A
0
value
• A
0
value / Temperature
• Remaining Time
Temperature / Remaining time
• A
0
value / Remaining time
• A
0
value / Remaining time / Temperature
If an alternative with more than one parameter is selected, the
parameters will be displayed alternately. The A
0
value always
remains at 0 until the thermal disinfection begins. When the
temperature reaches 85°C, the A0value will increase in line
with the temperature and time.
2.4.2 Display after program end
The display can be set to show one of the following alternatives
RQFHDSURJUDPKDV¿QLVKHG
• Max. temperature
• A
0
value
• A
0
value / Max temperature
If an alternative with more than one parameter is selected,
the parameters will be displayed alternately.