29
The Opera incorporates a battery discharge
indicator, situated on the rear side of the controls
console (see Fig. 1). The display shows ten levels of
battery state ranging from fully charged on the right
to very low on the left (green, through amber to red)
For more details of caring for your lifter battery
refer to the ‘Battery Care’ literature, ARJO Part No.
KDX01660.GB.
It is recommended that the battery is removed from
the lifter and recharged when the display reaches
the yellow range, but lifting is possible until the
display shows the red flashing light, at this point,
the battery must be recharged as soon as possible.
Recharging the battery pack before it reaches a low
state of battery charge or certainly totally
discharged will prolong its life.
Your lifter is fitted with an audible warning device
this will sound when the battery discharge indicator
reaches the red light range. The audible warning
will sound for approximately thirty seconds and
will start when a function button is pressed.
Pressing the emergency stop button will
temporarily silence this function, removing and
replacing the battery with one fully charged will
silence the alarm until low battery condition re-
activates it.
Danger: The charger is for indoor use
only.
Only use the charger unit in a dry
environment, do not use it in the
bathroom.
Do not expose the charger unit or battery
pack to rain or spray and do not immerse
in water.
Only use the ARJO battery that is
supplied to be used with the Opera
The battery charger is for use only with
ARJO supplied batteries that are to be
used with the Opera
The battery charger is for use with sealed
lead acid batteries only.
Under no circumstances should the
charger be used to attempt to recharge
non-rechargable batteries.
Do not attempt to open or tamper with the
charger unit in any way, for any repair the
charger must be sent to the manufacturer
The mains electricity socket must be
easily accessible. should a faulty
condition occur switch off and remove the
connection plug from the socket.
Only use ARJO components that have
been specifically designed for the purpose
when charging batteries.
Warning: To avoid overheating, the
charger must not be covered whilst in use.
No smoking or naked flames in battery
vicinity.
Do not expose the charger unit to dust.
Do NOT charge batteries in a sealed
container.
Do NOT place batteries near, or dispose
of, in a fire.
Do NOT short circuit a battery.
Warning: Do NOT store batteries at
temperatures in excess of 60°C (140°F).
Do NOT crush, puncture, open, dismantle
or otherwise mechanically interfere with
batteries.
Should the battery casing become
cracked, and electrolyte come into contact
with skin or clothing, wash immediately
with water.
If the electrolyte contacts the eyes, wash
immediately with copious amounts of
water, and seek medical attention.
When disposing of batteries, contact the
appropriate local authority for advice.
The abbreviation “Pb” shown adjacent to
the recycling and trash bin symbols on the
battery pack label is the element symbol
for lead, and indicates that the battery
contains lead and therefore should not be
disposed of in the normal manner but
must be recycled.
•Note: The battery discharge
indicator has an energy saving function
which automatically switches off the
display if a function button has not been
operated for at least 30 seconds. The
moment a button is pressed to operate
any function the display will re-start.