Care and Maintenance
Cleaning and Disinfection
To clean the Sara Stedy:
1) Use a cloth soaked with water to remove excessive residues.
2) Use a damp cloth with warm water mixed with a cleaning fluid.
3) Scrub the equipment from top to bottom to remove any deposit.
4) Use a clean damp cloth with clear water to wipe off all spots of cleaning fluids.
5) Wipe clean with dry cloth.
To disinfect the Sara Stedy, clean the equipment first, then wipe it using a solution containing
one of the compatible disinfectants below:
* Discolouration may occur over time if this agent is used.
Disinfection methods should comply with the local or national guidelines (Decontamination of
Medical Devices), depending on the Healthcare Facility or country of use.
If uncertain, consult local Infection Control specialist.
Sling Inspection and Care
See the documentation that comes with your sling.
Disinfectant Agents
Sara
Stedy
Pivoting Seat Halves
(seat pads)
Kneepad
Quaternary Ammonium ≤ 0.28% (2,800 ppm)
X X X
Isopropyl Alcohol ≤ 70%
X X X
Phenolic ≤ 1.56% (15,600 ppm)
X X X
Peracetic Acid ≤ 0.25% (2,500 ppm)
X X X
Bleach ≤ 1% (10,000 ppm)
X X* X*
Hydrogen Peroxide ≤ 6%
X X X
WARNING: To prevent cross-contamination, always follow the disinfection
instructions in this IFU.