Hydromassage (optional)
7
Hydrosound (optional)
Install the suction cover
before use.
Fill up with water to cover the
Hydromassage system outlet
(A) and top sensor (B).
Adjust the direction
of the water-jet.
Press the Hydrosound system
button. The default runtime is
15 minutes.
Risk of
entrapment
A
B
After treatment,
turn off the
Hydromassage
system.
After treatment, press the Hydromassage
system button to turn the Hydromassage
system off.
Never use
when pregnant.
Risk of
burns and
hearing
damage.
Do not touch the Transducer
Plate and keep patients
head above water.
Fill up with water to cover the
Hydrosound system Transducer Plate.
Add Hydrosoundwater conditioner.
Shower the patient
7
Open the drain. Raise the back rest. Let the
patient hold on to the hand rest.
Take the shower handle. Press the shower
button. Direct the shower handle away from
the patient.
Unlock the shower handle trigger if
lock button is activated.
Press the shower button.
Return the shower handle.
Let the patient feel the water by hand
before directing the water stream on the
patient. Adjust temperature if needed.
Risk of scalding
Shower the patient.
Press the trigger on the shower handle.
Check the water temperature with a bare
hand. Adjust temperature if needed.
Lower the bathtub to its lowest height.
+
-
After treatment,
turn off the
Hydrosound
system.
Adjust the
Hydrosound intensity.
20
P300
P220
End the bath
Dry the patient. Keep the patient warm using
towels or a bathrobe.
Lower the bathtub to its lowest height.
Risk of
pinching
Raise the lift hygiene chair enough to give space
between the patient’s genitals and the bathtub rim.
Help the patient to move the legs over the rim
while rotating the lift away from the bathtub.
Make sure there is enough space between
the bathtub side and lift hygiene chair.
Lower the lift hygiene chair.
Dry the remaining wet areas.
18
19
20
21
22 23 24 25
26
27
28
29
30
31 32
Press the Hydromassage system button.
The default runtime is 15 minutes.
Start with low intensity. Adjust if needed.
After each patient:
clean and disinfect the
bathtub according to
the instructions in the
IFU or Cleaning and
disinfection QRG.
33
Risk of contamination
ArjoHuntleigh AB
Hans Michelsensgatan 10
211 20 Malmö, Sweden
www.arjo.com
08.AR.08_1EN • 09/2020 • Page 2/2