Arjo Typhoon - Water Quality Requirements

Arjo Typhoon
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
Arjo recommends that water used in the pre-rinsing, washing and
¿QDOULQVLQJSKDVHVVKRXOGEHRIGULQNLQJTXDOLW\LQDFFRUGDQFH
with the guidelines. Recommended water quality is drinking water
ZLWKPD[G+'HWDLOHGLQIRUPDWLRQDERXWDFFHSWDEOHZDWHUTXDOLW\
can be found in “Guidelines for Drinking Water Quality, published
by WHO.
Follow local standards if they are stricter that Arjo’s
recommendations. Especially when treated water is used for the
¿QDOZDVKLQJGLVLQIHFWLRQSKDVH$W\SLFDOVSHFL¿FDWLRQIRUWUHDWHG
water is:
pH 5.5 to 8
Conductivity <30 μs.cm-1
TDS <40 mg/l
0D[LPXPKDUGQHVV <50 mg/l
Chlorine <10 mg/l
Heavy metals <10 mg/l
Phosphates <0.2 mg/l as P2O5
Silicates <0.2 mg/l as SiO2
(QGRWR[LQV <0.25 EU/ml
Total number of micro-
organisms
<100 per 100 ml
Additional advice should also be obtained from the manufacturers
of chemical and medical equipment.