2.8 Function check
Note! Make sure that the suction hose in the dosing pump
is full of agent before beginning. The dosing pumps can be
RSHUDWHGVHSDUDWHO\LQWKHVHUYLFHSURJUDPWR¿OOWKHKRVHV
ZLWKDJHQWOLQHVH7HFKQLFDOPDQXDO
&KHFNWKDWWKHÀXVKHUGLVLQIHFWRULVFRQQHFWHGWRWKHFRUUHFW
supply voltage and that it is protected by a fuse of the correct
rating as shown on the type plate.
Turn on the isolator switch. Check that all lamps on the panel
OLJKWXSDQGWKDWWKHEX]]HUVRXQGV7KHGLVSOD\WKHQVKRZV
the current program version, followed by the preselected
program. Open the water valves.
If the machine has detergent dosing: Press
and check
that the light in the button comes on.
Check that the temperature of the incoming water is correct.
7KHWHPSHUDWXUHPXVWEHLQWKH&UDQJHIRUKRWZDWHU
DQGQRWOHVVWKDQ&IRUFROGZDWHU
Note! Make sure that the suction hose in the dosing pump
is full of agent before beginning. The dosing pumps can be
RSHUDWHGVHSDUDWHO\LQWKHVHUYLFHSURJUDPWR¿OOWKHKRVHV
ZLWKDJHQWOLQHVH7HFKQLFDOPDQXDO