Electromagnetic Flowmeter M-910
40
Operation manual
1 0.0
2 0.00
3 0.000
4 0.0000
M910 confirms execution with string „Ok<cr>”.
Example:
Command „FFR3<cr>” sets flowrate resolution “0.000”. If query „FFR?<cr>” is sent, flowmeter returns response
in format „3<cr>”.
Volume resolution
FVR(?)<CPD> { 0 | 1 | 2 | 3 | 4 }
Following resolution can be set:
0 0
1 0.0
2 0.00
3 0.000
4 0.0000
M910 confirms execution with string „Ok<cr>”.
Example:
Command „FVR3<cr>” sets volume resolution “0.000”. If query „FVR?<cr>” is sent, flowmeter returns response
in format „3<cr>”.
Flowrate user unit
FFU(?)<CPD>
Command sets text for flowrate user unit.
<CPD>
It represents user units expressed as 5 ASCII characters. M910 confirms execution with string „Ok<cr>”. In case of
query M910 returns set user unit.
Example:
Command „FFU l/m <cr>” sets flowrate user unit l/m “. After query FFU?<cr>” flowmeter returns string l/m
<cr>”.
Volume user unit
FVU(?)<CPD>
Command sets text for volume user unit.
<CPD>
It represents user units expressed as 5 ASCII characters. M910 confirms execution with string „Ok<cr>”. In case of
query M910 returns set user unit.
Example:
Command „FVU dm3 <cr>” sets volume user unit dm3 “. After query „FVU?<cr>” flowmeter returns string
dm3 <cr>”.
Conversion constant for flowrate user unit
FFC(?)<DNPD>