Electromagnetic Flowmeter M-910
56
Operation manual
13 Terminology
Special symbols and terms.
Flowrates:
Q
1%
- minimum applicable flowrate (the least flowrate which has guaranteed measuring accuracy – depends on
diameter – see table 2 M910 flowrates).
Q
5%
- recommended minimum flowrate (least flowrate which has the best measuring accuracy depends on
diameter – see table 2 M910 flowrates).
Q
N%
- recommended nominal flowrate (nominal flowrate in which is flowmeter usually calibrated depends on
diameter – see table 2 M910 flowrates). You can predetermine this nominal flowrate in your order.
Q
50%
- recommended maximum flowrate (maximum flowrate which is usually used in industrial applications
depends on diameter – see table 2 M910 flowrates).
Q
100%
- maximum applicable flowrate (flowrate limit which has guaranteed measuring accuracy depends on
diameter – see table 2 M910 flowrates).
Q
MAX
- maximum applicable overload (Q
125%
) (maximum flowrate which can be still measured depends on
diameter – see table 2 M910 flowrates).
Abbreviations:
QI - current output constant. It represents flowrate for current 20 mA.
QF - frequency output constant. It represents flowrate for frequency 1000 Hz.
QP - impulse output constant. It represents volume for 1 impulse.
QD - constant for dosing. It represents volume for 1 dose.
PF1 - flowrate limit constant. It represents low limit flowrate. Crossing this limit activates the appropriate
digital output.
PF2 - flowrate limit constant. It represents high limit flowrate. Crossing this limit activates the appropriate
digital output.
H - flowrate limit constant. It represents hysteresis by evaluating limits PF1 and PF2.
Auxiliary volume counter second Total Volume counter. Can be cleared by pushing “RIGHT” key. It is usually
used for measuring volume during day, month etc.
RS232 serial bus. It enables remote control of instruments by a computer. Only one instrument can be connected
to one RS232 bus. Cable length between PC and instrument is limited to app. 10 metres.
RS485 serial bus. It enables remote control of instruments by a computer. To the RS485 can be connected more
instruments (max. 16). Total cable length is limited to app. 800 metres.