IEC 60601-1-2 Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The ARTHREX HD3Video System is intended for use in the electromagnetic environment specified below. The customer or user of the
ARTHREX HD3 Video System should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance Level
Electromagnetic environment - guidance
Electrostatic discharge [ESD]
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered by synthetic material the relative
humidity should be at least 30%.
Electrical fast transient / burst
IEC 61000-4-4
± 2 kV differential mode
± 1 kV for input / output
lines
100 khz Cycling
Frequency
± 2 kV differential mode
± 1 kV for input / output
lines
100 khz Cycling
Frequency
Mains power should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode.
± 2 kV common mode
± 1 kV differential
mode.
± 2 kV common mode
Mains power should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
U
T
= 0%,0.5 cycle
(0, 45, 90,135, 180,
225, 270, and 315
0
)
U
T
= 0%;1 cycle
U
T
= 70%;25/30 cycles
(@0 degrees)
U
T
= 70%; 250/300
cycles
U
T
= 0%,0.5 cycle
(0, 45, 90,135, 180,
225, 270, and 315
0
)
U
T
= 0%;1 cycle
U
T
= 70%;25/30 cycles
(@0 degrees)
U
T
= 70%; 250/300
cycles
Mains power should be that of a typical commercial
or hospital environment. If the user of the
ARTHREX HD3 Video System requires continued
operation during power mains interruptions, it is
recommended that the ARTHREX HD3 Video
System be powered from an uninterruptible power
supply or a battery.
Power Frequency [50/60 Hz]
magnetic field.
IEC 61000-4-8
3 A/m
3 A/m @50 & 60Hz
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.