EasyManuals Logo

Associated Research Hypot II 4500D User Manual

Default Icon
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
SPECIFICATIONS
14
GROUND
CONTINUITY
CHECK
Current : DC 0.1 A ± 0.01A, fixed
Max. ground resistance : 1 ± 0.1, fixed
GENERAL SPECIFICATIONS
Remote Control and
Signal Output
The following input and output signals are provided through two
9 pin D type connectors;
1. Remote control: test, reset, interlock
2. Outputs: pass, fail, test in process
Program Memory 5 sets
Security Lockout capability to avoid unauthorized access to test set-up
program.
Line Cord Detachable 7ft. (2.13m) power cable terminated in a three prong
grounding plug.
Terminations 5ft. (1.52m) high voltage and return leads (2) with clips and a
standard U.S. style (NEMA 5-15) remote receptacle box for
testing items terminated with a line cord. International
receptacles also available.
Mechanical Tilt-up front feet
Dimensions (W x H x D) 17 x 5.8 x 12 in. (432 x 147 x 305mm)
Weight: 56.70 Lbs. (25.92 kgs)
Environmental
Operating Temperatures: 32°-104°F (0°-40°C)
Relative Humidity : 0 to 80%
Calibration Traceable to National Institute of Standards and Technology
(NIST). Calibration controlled by software. Adjustments are
made through front panel keypad in a restricted access calibration
mode. Calibration information stored in non-volatile memory.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research Hypot II 4500D and is the answer not in the manual?

Associated Research Hypot II 4500D Specifications

General IconGeneral
BrandAssociated Research
ModelHypot II 4500D
CategoryTest Equipment
LanguageEnglish

Related product manuals