QUICK START
23
QUICKSTART
This quick start guide assumes the operator has some familiarity with automated hipot
testing and desires to use the “default” settings on the instrument. The default settings
shown will remain in memory unless you choose to override them with your own test
program. The default settings will appear in memories 1-5. The instrument default settings
are as follows:
DEFAULTS
• Input Voltage: 115 or 230 volts AC country specific
(rear panel switch selectable)
• Lock: OFF
• Voltage Output: 1,240 volts AC
• High Current Trip: HI-LMT: 10 mA
• Low Current Trip: LO-LMT: 0 mA
• Ramp Timer: 0 second
• Dwell Timer: 1 second
• AC Output Frequency: 60Hz
• Continuity: OFF