EasyManuals Logo

ATAGO PALETTE Series User Manual

ATAGO PALETTE Series
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
7. Measurement on sample
-9-
●When measuring a substance harmful to the human body, be well aware of its properties
and put on protective gloves, mask, etc.
●Do not use any metallic implement for sampling because the metal tool possibly damages
the prism surface.
●Do not apply water or sample over any part other than the surface of the prism.
Application on any other part of the instrument may result in a malfunction.
CAUTION
WARNING
(1)Clean the prism surface thoroughly.
(2)Apply a sample.
(3)Press the START/OFF button.
(4)Measurement value is displayed.
(5)Wipe off the sample.
  Wipe off the sample, and wipe and clean the Prism and the Sample Stage with tissue paper
moistened with water.
*
If [nnn] is displayed, covering the sample stage by hand, press the START/OFF button again.
START
OFF
ZERO
START
OFF
ZERO
START
OFF
ZERO
20.4
1 2.3
Completely cover prism surface
   (Min. 0.1mL)

Questions and Answers: