FG_D6_PROFIBUS_EN_01 / 2020-06-17
Parameters
Downloading of the parameters
All the parameters values below have a tratment by the ATEQ device as Long format with fixed comma
(10
-3
). A Long is a two words set.
Identifier N°
Meaning
*ERD Only
Value
Dec
Hexa
1
0001
“FILL TIME”
Fill time
0 > 650 seconds
2
0002
“STAB TIME”:
Stabilization time
0 > 650 seconds
3
0003
“TEST TIME”
Test time
0 > 650 seconds
4
0004
“P.UP TIME” *
Rise time
0 > 650 seconds
5
0005
“P.DOWN TIME” *
Drop time
0 > 650 seconds
9
0009
“DUMP TIME”
Dump time
0 > 650 seconds
10
000A
“COUPL. A”:
Coupling time 1
0 > 650 seconds
11
000B
“COUPL. B”:
Coupling time 2
0 > 650 seconds
21
0015
“TYPE”:
Test type
Invalid
Direct flow
Indirect
Operator
Direct ERD *
Indirect ERD *
Quick test
0000
1000
2000
3000
4000
5000
6000
29
001D
“Inter-Cycle”:
Time between 2 chained cycles
0 > 650 seconds
48
0030
“DURATION”
Maintain time of the result during stamp
0 > 650 seconds
50
0032
“Min Press”
Minimum pressure value
- 9999 > 9999
51
0033
“Max Press”
Maximum pressure value
- 9999 > 9999
53
0035
“Press. UNIT” Pressure unit.
Refer to the “Unit” table.
60
003C
“Max Flow”
Maximum reject value
- 9999 > 9999
61
003D
“Max Rework”
Maximum reject value in recovery
- 9999 > 9999
62
003E
“Min Flow”
Minimum reject value
- 9999 > 9999
63
003F
“Min Rework”
Minimum reject value in recovery
- 9999 > 9999