EasyManua.ls Logo

AVL DITEST COUNTER - Page 40

AVL DITEST COUNTER
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Errors and Error Elimination
AVL DiTEST Counter
User manual
36
31009
The exhaust probe is blocked.
-
Check the exhaust
probe.
31204
The calibration is expired.
-
Schedule a
calibration
appointment
immediately. A
measurement can
only be performed
after a new
calibration.
31501
Error when accessing the SD
card.
The SD card is defective.
Contact the service
hotline.
31502
No SD card found.
There is no SD card in the
reader.
Contact the service
hotline.
31601
Connect correct heating hose.
Either an incorrect or defective
heating hose is connected.
Connect a suitable
heating hose.
40209
The maintenance cover is
open.
-
Close the
maintenance cover.
40401
Realtime clock unreliable.
-
Contact the service
hotline.
40601
Pump error
The pump is defective.
Contact the service
hotline.
40801
Heating error hose.
The NTC sensor of the heating
hose is defective.
Contact the service
hotline.
40802
Heating error VPR.
The NTC sensor of the VPR is
defective.
Contact the service
hotline.
40803
Heating error Corona.
The NTC sensor of the Corona
is defective.
Contact the service
hotline.
40804
Heating error FCEM.
The NTC sensor of the FCEM
is defective.
Contact the service
hotline.
40805
Heating error CFO.
The NTC sensor of the CFO is
defective.
Contact the service
hotline.
40806
NTC error mainboard.
The NTC sensor on the
mainboard is defective.
Contact the service
hotline.
40807
NTC error electrometer.
The NTC sensor on the
electrometer board is defective.
Contact the service
hotline.
40808
RHT error ambient.
The ambient rH/T sensor is
defective.
Contact the service
hotline.
40809
RHT error CFO.
The CFO rH/T sensor is
defective.
Contact the service
hotline.
40810
RHT error Corona.
The Corona rH/T sensor is
defective.
Contact the service
hotline.
40811
Internal ADC Error
The ADC is defective.
Contact the service
hotline.
41204
A calibration checksum
mismatch.
-
A measurement can
only be performed
after a new
calibration.

Related product manuals