Errors and Error Elimination
The exhaust probe is blocked.
The calibration is expired.
Schedule a
calibration
appointment
immediately. A
measurement can
only be performed
after a new
calibration.
Error when accessing the SD
card.
The SD card is defective.
Contact the service
hotline.
There is no SD card in the
reader.
Contact the service
hotline.
Connect correct heating hose.
Either an incorrect or defective
heating hose is connected.
Connect a suitable
heating hose.
The maintenance cover is
open.
Close the
maintenance cover.
Realtime clock unreliable.
Contact the service
hotline.
Contact the service
hotline.
The NTC sensor of the heating
hose is defective.
Contact the service
hotline.
The NTC sensor of the VPR is
defective.
Contact the service
hotline.
The NTC sensor of the Corona
is defective.
Contact the service
hotline.
The NTC sensor of the FCEM
is defective.
Contact the service
hotline.
The NTC sensor of the CFO is
defective.
Contact the service
hotline.
The NTC sensor on the
mainboard is defective.
Contact the service
hotline.
The NTC sensor on the
electrometer board is defective.
Contact the service
hotline.
The ambient rH/T sensor is
defective.
Contact the service
hotline.
The CFO rH/T sensor is
defective.
Contact the service
hotline.
The Corona rH/T sensor is
defective.
Contact the service
hotline.
Contact the service
hotline.
A calibration checksum
mismatch.
A measurement can
only be performed
after a new
calibration.