68
REPLACE FLUID PACK
When the monitored fluid level in the ISE SnapPak
TM
reaches 5% remaining,
the analyzer will print REPLACE FLUID PACK at the end of each sample
report. This message is printed only. Replace the ISE SnapPak
TM
following
the instructions outlined in the Operator Manual.
To maximize the life of the ISE SnapPak
TM
, continue running the analyzer until
a STANDARD A NOT DETECTED message is displayed. This will
ensure that the ISE SnapPak
TM
is completely depleted before replacement.
CHECK TEMPERATURE (9140/9180/9181 only)
On the 9140/80/81 models, the analyzer will display CHECK
TEMPERATURE at the end of each measurement, if the temperature
measured is out of range (10.0°C to 40°C). The temperature sensor is located
in the right side electrode holder.
Possible causes and remedies:
• Ensure that the sample sensor cable is securely plugged in.
• Check whether room temperature is within specified limits
(15°C to 32°C / 60°F to 90°F).
• Perform temperature adjustment procedure.
• Replace SBC board or right side electrode holder.
VALVE OVERTEMP! CHECK VALVES
In case one of the valve solenoids overheats, the message VALVE
OVERTEMP! CHECK VALVES will be displayed. Each solenoid
incorporates a thermo-fuse, which will reset automatically.
Possible causes and remedies:
• Turn power off for at least one hour.
• Check fuse F3 (24V).
• Check whether room temperature is within specified limits
(15°C to 32°C / 60°F to 90°F).